Vaporizer NEW 300 SERIES

Placeholder image

    Use only for Isoflurane
    Contact Sales for Fee

3 Year Warranty

Weight/Dimensions: (Screw fill, cagemount)

Weight: 6.8kg

Height: 183mm

Depth: 140mm

Width: 135mm

Features: Calibrated specifically for the specified anesthetic drug, Calibrated using a refractometer, Temperature, flow & duration of use compensation,Safety lock to help prevent accidental turn-on of dial

Specifications: Calibrated at 21°C and 35°C Calibration range: 0.25%, 0.5% through to 5% v.v. in 0.5% increment, Fluctuations in temperature, flow rate & duration of use is automatically compensated for during use,When used with I.P.P.V the variation in output is negligible

Resistance to Gas Flow: 5cm wg. at "OFF" setting at 5 liter/min oxygen, 21-29cm wg. when delivering vapor in 5 liter/min oxygen at 21°C

Liquid Capacity: Amount of anesthetic drug to fully charge the vaporizer - 125ml, Amount retained by the wick system - 35ml

Normal Range: Fresh gas flow - 200cc to 15 liters/min, Ambient temperature - 18°C to 35°C

Options: Selectatec® mounting system that allows optimum flexibility of use, Screw filling, Cage mount (23mm male/female tapers) mounting, Isoflurane, Halothane, Sevoflurane, Enflurane,*Keyed filling for safer and pollution reduced filling and draining


Placeholder image

Vaporizer OHIO

Placeholder image

    Use only for Isoflurane
    Contact Sales for Fee

1Year Warranty with annual service

Weight/Dimensions: (Screw fill, cagemount)

Weight: 6.8kg

Height: 183mm

Depth: 140mm

Width: 135mm

Features: Calibrated specifically for the specified anesthetic drug, Calibrated using a refractometer, Temperature, flow & duration of use compensation,Safety lock to help prevent accidental turn-on of dial

Specifications: Calibrated at 21°C and 35°C Calibration range: 0.25%, 0.5% through to 5% v.v. in 0.5% increment, Fluctuations in temperature, flow rate & duration of use is automatically compensated for during use,When used with I.P.P.V the variation in output is negligible

Resistance to Gas Flow: 5cm wg. at "OFF" setting at 5 liter/min oxygen, 21-29cm wg. when delivering vapor in 5 liter/min oxygen at 21°C

Liquid Capacity: Amount of anesthetic drug to fully charge the vaporizer - 125ml, Amount retained by the wick system - 35ml

Normal Range: Fresh gas flow - 200cc to 15 liters/min, Ambient temperature - 18°C to 35°C

Options: Selectatec® mounting system that allows optimum flexibility of use, Screw filling, Cage mount (23mm male/female tapers) mounting, Isoflurane, Halothane, Sevoflurane, Enflurane,*Keyed filling for safer and pollution reduced filling and draining


Placeholder image

Vaporizer ISOTEC 4

Placeholder image

    Use only for Isoflurane
    Contact Sales for Fee

1Year Warranty with annual service

Weight/Dimensions: (Screw fill, cagemount)

Weight: 6.8kg

Weight: 6.8kg

Weight: 6.8kg

Weight: 6.8kg

Features: Calibrated specifically for the specified anesthetic drug, Calibrated using a refractometer, Temperature, flow & duration of use compensation,Safety lock to help prevent accidental turn-on of dial

Specifications: Calibrated at 21°C and 35°C Calibration range: 0.25%, 0.5% through to 5% v.v. in 0.5% increment, Fluctuations in temperature, flow rate & duration of use is automatically compensated for during use,When used with I.P.P.V the variation in output is negligible

Resistance to Gas Flow: 5cm wg. at "OFF" setting at 5 liter/min oxygen, 21-29cm wg. when delivering vapor in 5 liter/min oxygen at 21°C

Liquid Capacity: Amount of anesthetic drug to fully charge the vaporizer - 125ml, Amount retained by the wick system - 35ml

Normal Range: Fresh gas flow - 200cc to 15 liters/min, Ambient temperature - 18°C to 35°C

Options: Selectatec® mounting system that allows optimum flexibility of use, Screw filling, Cage mount (23mm male/female tapers) mounting, Isoflurane, Halothane, Sevoflurane, Enflurane,*Keyed filling for safer and pollution reduced filling and draining


Placeholder image

Vaporizer ISOTEC 5

Placeholder image

    Use only for Isoflurane
    Contact Sales for Fee

1 Year Warranty

Weight/Dimensions: (Screw fill, cagemount)

Weight: 7.0kg

Height: 250mm

Depth: 164mm

Width: 110mm

Specifications: Calibrated at 21C using an oxygen carrier gas at a flow of 5 litres/minute and they are temperature, flow and pressure compensated within the specific operating range.

Performance: Accuracy at 5 litres/minute20@ 21˚C + 2˚C
Vaporizer 5% = + 0.25% of delivered agent or + 15% of delivered setting (whichever is greater)
Vaporizer 8% = + 0.4% of delivered agent or + 15% of delivered setting (whichever is greater)

Resistance to Gas Flow:Out of Circuit Resistance = 10-15cm wg / Vaporizer Setting = On + 0% Flow Rate = 5 litres/minute, Oxygen / Temperature = 21 + 2˚C

Liquid Capacity: Amount of liquid to fully charge a vaporizer with dry wicks = 300 mililitres (nominial) The amount retained by the wick system = 75 mililitres (nominial)
Vaporizer 5% - The amount of liquid to fill the vaporizer from the minimum to the maximum mark = 170 mililitres (nominial)
Vaporizer 8% (Servotec) - The amount of liquid to fill the vaporizer from the minimum to the maximum mark = 137 mililitres (nominial)

Enviromental Conditions: Operating Temperature Range = 16˚ to 35˚C Storage Temperature Range = 40˚ to 65˚C Note: The vaporizer packaging should be protected from condensation


Placeholder image

Vaporizer SERVOTEC 5

Placeholder image

    Use only for Isoflurane
    Contact Sales for Fee

1 Year Warranty

Weight/Dimensions: (Screw fill, cagemount)

Weight: 6.8kg

Height: 183mm

Depth: 140mm

Width: 135mm

Features: Calibrated specifically for the specified anesthetic drug, Calibrated using a refractometer, Temperature, flow & duration of use compensation,Safety lock to help prevent accidental turn-on of dial

Specifications: Calibrated at 21°C and 35°C Calibration range: 0.25%, 0.5% through to 5% v.v. in 0.5% increment, Fluctuations in temperature, flow rate & duration of use is automatically compensated for during use,When used with I.P.P.V the variation in output is negligible

Resistance to Gas Flow: Resistance to Gas Flow: 5cm wg. at "OFF" setting at 5 liter/min oxygen, 21-29cm wg. when delivering vapor in 5 liter/min oxygen at 21°C

Normal Range: Fresh gas flow - 200cc to 15 liters/min, Ambient temperature - 18°C to 35°C

Liquid Capacity: Amount of anesthetic drug to fully charge the vaporizer - 300ml, Amount retained by the wick system - 35ml

Options: Selectatec® mounting system that allows optimum flexibility of use, Screw filling, Cage mount (23mm male/female tapers) mounting, Isoflurane, Halothane, Sevoflurane, Enflurane,*Keyed filling for safer and pollution reduced filling and draining


Placeholder image

Enhanced NEW Tec 3

Placeholder image

   Use only for Isoflurane / holds 250ml
    Contact Sales for Fee

Updated Information is coming soon

Weight/Dimensions: (Screw fill, cagemount)

Weight: 7.0kg

Height: 250mm

Depth: 164mm

Width: 110mm

Specifications: Calibrated at 21C using an oxygen carrier gas at a flow of 5 litres/minute and they are temperature, flow and pressure compensated within the specific operating range.

Performance: Accuracy at 5 litres/minute20@ 21˚C + 2˚C
Vaporizer 5% = + 0.25% of delivered agent or + 15% of delivered setting (whichever is greater)
Vaporizer 8% = + 0.4% of delivered agent or + 15% of delivered setting (whichever is greater)

Resistance to Gas Flow:Out of Circuit Resistance = 10-15cm wg / Vaporizer Setting = On + 0% Flow Rate = 5 litres/minute, Oxygen / Temperature = 21 + 2˚C

Liquid Capacity: Amount of liquid to fully charge a vaporizer with dry wicks = 300 mililitres (nominial) The amount retained by the wick system = 75 mililitres (nominial)
Vaporizer 5% - The amount of liquid to fill the vaporizer from the minimum to the maximum mark = 170 mililitres (nominial)
Vaporizer 8% (Servotec) - The amount of liquid to fill the vaporizer from the minimum to the maximum mark = 137 mililitres (nominial)

Enviromental Conditions: Operating Temperature Range = 16˚ to 35˚C Storage Temperature Range = 40˚ to 65˚C Note: The vaporizer packaging should be protected from condensation